English
28
Cleaning
Remove the AC adapter from the device when cleaning the device.
When the main body or cuff is dirty, wipe them fully by using a gauze or cloth
dampened with warm water and a neutral detergent avoiding excess water.
Do not use the moisten cloth etc. to wipe the DC jack and air socket. The
DC jack and air socket must remain dry.
To prevent a risk due to infection, disinfect the main body and cuff
regulary. When disinfecting them, wipe them gently by using the gauze or
dampened cloth with local antiseptic solution then wipe the moisture off
the surface by using a dry soft cloth.
Use the following disinfectants to clean the main body and cuff.
Ethanol (70%)
Isopropanol (70%)
Chlorhexidine Gluconate Solution (0.5%)
Benzalkonium Chloride Solution (0.05%)
Sodium Hypochlorite (0.05%)
Clean the device about once every month, basing on a policy or
instruction specified in the hospital or clinic.
CAUTION
The blood pressure monitor is not waterproof device. Do not splash water
on it and avoid exposure to moisture.
Do not use a organic solvent such as thinner or benzine.
The blood pressure monitor can not be sterilized by autoclave, EOG or
formaline gas etc.
Regular inspection
The blood pressure monitor is a precision device. Therefore, inspect it
regularly. Request an inspection to the dealer where you have purchased
the device when the device is in needs of an inspection,
The cuff is consumable. Regularly exchange the cuff with new one.