The measurement
is started.
Auscultation
setting mode
Model UM-211 is designed to detect the pulse and to inflate the cuff to a
systolic pressure level automatically.
If re-inflation occurs repeatedly, use the following methods.
Measurement with the SET Pressure
During the blood pressure measurement, re-inflation may occur.
A fixed pressure value can be set to avoid re-inflation.
1. Press the MODE button to go to the
pressurization value setting mode. The
current setting blinks.
2. Press
the or button to select a pressure
value about 30 mmHg or more above your
expected systolic pressure from the following.
AUTO : Automatic pressurization (default value)
220 : Pressure value of 220 mmHg (fixed)
250 : Pressure value of 250 mmHg (fixed)
280 : Pressure value of 280 mmHg (fixed)
3. Press the MODE button to go to the
auscultation setting mode.
Press the START/STOP button to start the
measurement. The device will proceed to
standby mode automatically when no
operation is made for a regular time.
The next measurement will be performed
with the new pressure value.
or
START
STOP