![AND UM-211 Instruction Manual Download Page 24](http://html.mh-extra.com/html/and/um-211/um-211_instruction-manual_2941387024.webp)
English
23
Auscultation measurement
The auscultation measurement is performed when the auscultation setting
mode is set to ON. Also, Press the START/STOP button while pressing
the MODE button to perform the auscultation measurement.
The auscultation measurement is returned to OFF automatically when the
device goes into standby mode.
1. Press the START/STOP button to start
pressurization. When conditions for the
pressurization to be completed will be
arranged, the device starts the constant
speed exhaustion after completing
pressurization.
2. The device exhausts at constant speed.
Press the MODE button to confirm the
systolic pressure value. Press the MODE
button again to confirm the diastolic
pressure value, and the device exhausts at
quick speed.
3. Press the button during exhausting at
constant speed to perform the additional
pressurization while the button is being
pressed. The additional pressurization
mark illuminates in order from bottom
during the additional pressurization. When
additional pressurization is applied up to
the systolic pressure value or more, the
systolic pressure value is cleared.
Note: When the device is pressured at
300mmHg or more, the device
performs forced exhaust
automatically.
A mark for the additional pressurization
4. Press the START/STOP button after
measuring to carry out the auscultation
measurement again.
Note: Allow at least three minutes
between measurements on the
same person.
START
STOP
Measurement
in progress
Pressurizing
Zero display
Starts inflation
The value is
fixed.
Decreasing by
exhaust
The diastolic
pressure value is
also confirmed,
and the
measurement
completes.