Operation
Standby mode
- The device goes into standby mode when the power is turned on, and a
current room temperature is displayed at the display for diastolic pressure.
- The device proceeds to standby mode when the START/STOP button is
pressed and held, or no operation is made for a regular time at all status
other than blood pressure mode and auscultation mode.
- Press the
or
button to read out the memory.
- Press the MODE button to proceed to the pressurization value setting
mode.
- Press and hold the MODE button to proceed to the clock setting mode.
- Press the START/STOP button to start the measurement.
Measurement standby mode
- The device proceeds to measurement standby mode when the auscultation
mode is set to OFF at the auscultation setting mode, or the MODE button is
pressed at the auscultation exhaust speed changing mode, or the
measurement is stopped.
- Also, the device proceeds to measurement standby mode when the
measurement is completed. In this case, the device remains measurement
results displayed.
- Press the
or
button to read out the memory.
- Press the MODE button to proceed to the pressurization value setting
mode.
- The device proceeds to the standby mode automatically after a regular
time.
- Press the START/STOP button to start the measurement.
Standby mode Measurement standby mode
The display defers
depending on a
setting.
A current temperature
is displayed.