29
15.6. Parameter List
Class
Item
Parameter
Content/use
func
Automatic power off function
poff
0
OFF
1
ON (Turns off after 5 minutes)
2
ON (Turns off after 10 minutes)
3
ON (Turns off after 15 minutes)
4
ON (Turns off after 30 minutes)
5
ON (Turns off after 60 minutes)
Automatic power on function
p-on
0
OFF
1
ON (The scale turns on automatically if the power
plug is connected.)
Display resolution
re5o
0
Low resolution (1/3,000)
1
Normal resolution (1/6,000 to 1/7,500)
2
High resolution (1/12,000 to 1/15,000)
3
Max resolution (1/30,000)
Zero tracking function
trc
0
OFF
1
ON
Stability band width
5t-b
0
±0.5d (width 1d)
1
±1.0d (width 2d)
2
±2.0d (width 4d)
Stability band time
5t-t
0
0.2 seconds
1
0.5 seconds
2
1.0 seconds
Response
Cond
0
Fast response / Poor vibration tolerance (for stable
environments)
1
↑↓
2
↑↓
3
↑↓
4
Slow response/stabilized display (prioritizes
stability)
:
factory settings
d: readability (scale interval or "division")
Automatic power off: When the specified interval has passed while the display is zero, the display is
stable and there is no key operation, the display turns off.
Display resolution: Changes the value for readability “d”. (Resolution = Readability / Weighing capacity)
Stability band width: When the weighed value is within a certain width (stability band width) and a
certain amount of time (stability band time) has passed, the scale judges that
weighing is stable and lights up the stability indicator. To ensure maximum
stability, set a large value. For faster judgment of stability at the expense of
precision, use a smaller value.
Stability band time: When the weighed value is within a certain width (stability band width) and a
certain amount of time (stability band time) has passed, the scale judges that
weighing is stable and the stability indicator lights up. To ensure maximum
stability, set a large value. For faster judgment of stability at the expense of
precision, use a smaller value.