14
7. CALIBRATION
This function adjusts the balance for accurate weighing.
Perform calibration in the following cases.
When the balance is first used.
When the balance has been moved.
For regular calibration.
7-1. Calibration using a weight
Prepare a calibration weight (sold separately) in advance.
1. Warm up the balance for at least half an hour with
nothing on the pan.
2. Press and hold the calibration (CAL) switch until
Cal
appears, and release the switch.
3. The balance displays
Cal 0
.
To change the calibration weight value, proceed to
step 4.
To use the calibration weight value in the balance
memory, proceed to step 5.
4. Press the SAMPLE key. The display shows the
calibration weight value in “grams” that is stored in
the balance. Use the following keys to change the
value.
SAMPLE key
To select the digit blinking to be
changed.
RE-ZERO key To set the value of the digit
selected.
PRINT key
To store the value and return to
step 3.
UNITS key
To cancel the value and return to
step 3.
Cal
Cal 0
06000.0
Cal 0
Set the weight using
the relevant keys
Release the CAL switch
Calibration (CAL) switch
Calibration (CAL) switch cover
Press and hold
the CAL switch