28
Change the function table
Selecting a way to set the density of a liquid
Select the liquid density input method from the function table below. The function table
is available only when the density measurement mode is selected.
Class Item
Param-
eter
Description
0
Water temperature
func
ldin
Liquid density input
1
Liquid density
The way to input liquid density.
Factory setting
Entering the density of a liquid (
ldin
=
0
)
1. Press the UNITS key to select SG .
2. Press and hold the UNITS key to display the
liquid density input mode. The display will show
the water temperature currently set (factory
setting: 25 °C).
3. Using the RE-ZERO (to increment the value) and SAMPLE keys (to shift the
selected digit), set the value and press the PRINT key to store.
To cancel the setting procedure and return to the density measuring mode, press
the UNITS key. The input value is not stored.
The relation between the water temperature and density is shown below.
°C
+0 +1 +2 +3 +4 +5 +6 +7 +8 +9
0 0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978
10 0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
20 0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
30 0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
40 0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
50 0.98804 0.98758 0.98712 0.98665 0.98618 0.98570 0.98521 0.98471 0.98422 0.98371
60 0.98320 0.98268 0.98216 0.98163 0.98110 0.98055 0.98001 0.97946 0.97890 0.97834
70 0.97777 0.97720 0.97662 0.97603 0.97544 0.97485 0.97425 0.97364 0.97303 0.97242
80 0.97180 0.97117 0.97054 0.96991 0.96927 0.96862 0.96797 0.96731 0.96665 0.96600
90 0.96532 0.96465 0.96397 0.96328 0.96259 0.96190 0.96120 0.96050 0.95979 0.95906
Entering the density of a liquid directly (
ldin
=
1
)
1. Press the UNITS key to select SG .
2. Press and hold the UNITS key to display the
liquid density input mode. The display will show
the liquid density currently set (factory setting:
1.0000 g/cm
3
)
3. Using the RE-ZERO (to increment the value) and SAMPLE keys (to shift the
selected digit), set the value and press the PRINT key to store.
To cancel the setting procedure and return to the density measuring mode, press
the UNITS key. The input value is not stored.
t
---
25
SG
d1.0000
SG
SG indicator
0.00
SG
HI