![AND AD-4406 OP-04 Instruction Manual Download Page 47](http://html1.mh-extra.com/html/and/ad-4406-op-04/ad-4406-op-04_instruction-manual_2941308047.webp)
AD-4406 Weighing Indicator
Page
45
13.2. Data Format
There are two types of data format set at F-Function F34 and F35.
The initial data format of F34 is shown below.
Header 1 F34, 35 : 9 (Status)
ST
Stable weight data.
US
Unstable weight data.
OL
Overload (Out of range).
HD
Hold weight data (The response of the “RW” and “RW,n” command).
Header 2 F34, 35 : E (Weight type)
GS
or
G
Gross data.
NT
or
N
Net data
TR
or
T
Tare data.
Data
F34, 35 : A, B, C, D (Weight data)
The first of the data bits is the polarity, "+" or "-".
When data is zero, the polarity is "+".
8 digits (bits) including the polarity and a decimal point (dot or comma).
In case of "Out of range", the data bits are replaced by spaces except for the
decimal point.
Unit
F34, 35 : F (Unit)
In case of CF07=0, the unit length is 2 digits.
In case of CF07=1, the unit length is 3 digits. Depending on the circumstances, an
A&D printer may not work correctly.
13.3. Command Format
Description of Command
When performing a command, the received command or reply data is sent back.
When the received command can not be performed such as the "busy" state, the
code "
I
" is sent back. In this case, insert a delay time. Electrical noise may be
the cause of this error.
13.2. Data Format
Carriage return (0Dh)
CF06=0,1 CF06=2
CF07=0 CF07=1
2 digits 3 digits
Header 1 Header 2
Unit
8-digit data
Terminator
T
S , S
G
+
,
0
0
0
0 0 0
.
g
k
T
S
S
U
L
O
D
H
G
N
T
S
G
T
N
R
T
g
t
g
k
g
t
g
k
l
t
b
l
l
t
b
l
z
o
y
c
z
o
y
c
Space code (20h)
Line feed (0Ah)
Decimal point :
Dot (2Eh) or
Comma (2Ch)