AD-4406 Weighing Indicator
Page
13
5.2.4. Specifying the Weighing Range of the First Range
Step 3 After displaying CAP for 2 seconds, single range or the weighing capacity will be
displayed. When dual range is used, CAP1 is displayed for 2 seconds. Triangle 1
will be displayed. Specify the range or the capacity with the and keys.
Press the ENTER key to store the parameter and proceed to the next step.
When pressing the ESC key, regardless of what is displayed, the indicator will
proceed to the next step without changing the parameter. The next step is zero
calibration in single range or specifying the second range resolution in dual
range.
5.2.5. Specifying the Second Range Resolution
Step 4 After displaying rAnGE2 for 2 seconds, the resolution with decimal point and
triangle 2 will be displayed. Specify the second range resolution in the same way
as the first range. The decimal point cannot be moved. Specify the second range
resolution greater than the first range.
Press the ENTER key to store the parameter and proceed to the next step.
When pressing the ESC key, regardless of what is displayed, the indicator will
proceed to the next step without changing the parameter.
5.2.6. Specifying the Second Range Capacity
Step 5 After displaying CAP2 for 2 seconds, the capacity with unit and decimal point will be
displayed. Specify the capacity in the same way as the first range. The capactity
should be greater than the first range.
Press the ENTER key to store the parameter and proceed to zero calibration.
When pressing the ESC key, regardless of what is displayed, the indicator will
proceed to zero calibration without changing the parameter.
5.2.7. To Get Stabilized Data
Step 6 Maintain the following conditions to calibrate the scale (indicator) correctly.
Maintain a constant temperature, stable power and stable input voltage from the
load cell.
Avoid direct sunshine or near the outlet of an air conditioner.
Do not install the scale (indicator) where there is a strong magnetic field.
Step 7 Turn the display on and leave it for several minutes.
5.2. Calibration Procedure