![AND AD-4406 OP-04 Instruction Manual Download Page 29](http://html1.mh-extra.com/html/and/ad-4406-op-04/ad-4406-op-04_instruction-manual_2941308029.webp)
AD-4406 Weighing Indicator
Page
27
Current Loop Output
Item
Parameter
Description
* 0
Displayed data
1 Gross
data
2 Net
data
3 Tare
data
F36
Output data
4
Gross data/Net data/Tare data
0 Stream
mode
1 Manual
mode
2
Auto print mode (+)
3
Auto print mode (+/-)
4
When accumulation, automatically output
F37
Output mode
* 5
No output
* 0
No delay
F38
Delay for continual data
1
2.0 seconds (F36=4, except F37=0)
0
600
bps
1 1200
bps
F39
Baud rate
* 2
2400 bps
bps: bit per second
*: Initial settings
Serial Interface
tem
Parameter
Description
* 0
Stream mode, command is
not
acceptable
1
Manual mode, command is effective
2
Auto print mode (+) , command is effective
3
Auto print mode (+/-), command is effective
4
When accumulation, automatically output
F40
Output mode
5 Command
mode
0 No
output
* 1
Manual, Fixed format
2
Auto, Fixed format
3
Manual, Format 1 (F34)
4
Auto, Format 1 (F34)
5
Manual, Format 2 (F35)
F41
Accumulated data output
at accumulated data
display
6
Auto, Format 2 (F35)
* 0
No delay
1 0.5
second
2 1.0
second
3 1.5
seconds
F42
Delay for continual data
4 2.0
seconds
* 0
Not used
F43
Command address
1
Used (Address selected at F06)
* 0
Approx. 1 second
F44
Time out
1 No
limitation
* 0
CR, LF
F45
Terminator
1 CR
*: Initial settings
6.2.
F-Functions