![AND AD-4406 OP-04 Instruction Manual Download Page 18](http://html1.mh-extra.com/html/and/ad-4406-op-04/ad-4406-op-04_instruction-manual_2941308018.webp)
Page
16
AD-4406 Weighing Indicator
Press and hold the ON/OFF key and press the ESC key.
No parameters are changed,
CAnCEL is displayed and the calibration mode is finished.
Step18 Press the ON/OFF key to turn the display off.
5.3. Weighing Range Function
The weighing range function can select "single range" or "dual range". Specify each
weighing interval (division) for the multi-interval instrument. Each weighing interval is
displayed according to a net value or gross value.
Note
When single range is used, performing this function is not required.
Example 1
The gross display.
Specified
parameters:
First range
Range = 50.00 kg, division 0.02 kg
Second range
Range = 100.00 kg (maximum capacity), division 0.1 kg
Display
0 kg to 50 kg :
The first range, division 0.02 kg.
50 kg to 100 kg :
The second range, division 0.1 kg.
division 0.02 kg
division 0.1 kg
0 kg 50 kg 100 kg Gross
Weighing value
Example 2
The net display using a 40kg tare value.
Specified parameters:
The same parameters as example 1.
Display
-40 kg to 50 kg :
The first range, division 0.02 kg.
50 kg to 60 kg :
The second range, division 0.1 kg.
division 0.02 kg
division 0.1 kg
-40 kg 0 kg 50 kg 60 kg Net
Tare value
90 kg 100 kg
40 kg Weighing value
5.3. Weighing Range Function