Page 36
10.1
Jobs to be done during an inspection
PARTS WORK
DONE
DESCRIPTION
PERIODS
CO SENSOR FILTER
Replaced
If present
1 YEAR
NO SENSOR FILTER
Replaced
If present
1 YEAR
THREE-WAY FILTER
Replaced
1 YEAR
WATER SEPARATOR SEAL
Replaced
Silicon seal
1 YEAR
SILICON HOSES
Replaced
All hoses
1 YEAR
INSULATION HOSE PROBE
Replaced
1 YEAR
CPU BATTERY
Replaced
RTC clock
3 YEARS
NO BATTERY
Replaced
2 YEARS
CALIBRATION OF ALL SENSORS AND
GENERAL FUNCTION CONTROL
Carried out
1 YEAR
SOFTWARE
Programmed
Last version
CLEANING
Carried out
1 YEAR
10.2
EMC conformity
Notes on EMC conformity:
The perturbing electromagnetic radiation emitted lies well below the legal limit value.
High voltage discharges on plastic parts (front panel) have no effects. Discharges on metal parts
(connections, probes) can initiate a restart (reset) of the device.
Strong high frequency fields (several watts) in the immediate vicinity of the device (~50cm) can lead to a
disturbance in measured values.
The source of the interference should be removed (e.g. mobile telephone).