3.6
Free fuels
In this menu, 8 fuels can be programmed.
Five fuels are pre-programmed, three more can be programmed.
For the USA: all 8 fuels are freely programmable.
Example:
⇒
Press the keys
and select the storage location with the arrow keys.
If no free fuel has yet been recorded, the following appears on the display:
Free fuels
6 ----
*: new M: Menu
If fuels have already been recorded, the following appears on the display:
Free fuels
6 samples
E: Edit P: Delete
*: Insert M: Menu
Select with
and
.
Press the key
to set up a new fuel
Fuel name
_
Specify the relevant name by entering text and then confirm with
this key
.
Calculation method
A2 B
Select the calculation method
A2, B or k with
and
. Continue with
When entering factors A2 and B, the exhaust gas losses are calculated according
to the recommendations for the measurement of exhaust gases from burners
for "extra light" heating oil or gas of the Swiss Agency for the Environment,
Forests and Landscape ("BUWAL") dated 15 August 1996. This is the usual
calculation method.
If the k factor is entered, the calculation of exhaust gas losses is carried out
according to Siegert.
Is a soot measurement necessary in the case of this fuel?
Soot measurement
Yes
Select with
and
. Continue with
If you have selected calculation method A2 B:
⇒
Enter the appropriate parameters for A2. With this key
you can delete entries.
Enter parameters
A2 _
Select with
-
,
and continue with
Sample entry : A2 = 0.68 type
.
⇒
Now enter the appropriate parameters for B
Page 21