NO
TG
<= 40 ppm
<= 50°C
If this display still persists after a number of attempts, please contact anapol.
9.4
Error message during measurement
Alarm CO! 1670 ppm
Please wait
Cause: you are measuring a combustion process with a CO content
of more than 1,600 ppm.
Solution: stop the measurement procedure immediately! These
exhaust gases can put the device out of commission for a long time!
You need a different kind of measuring device for these exhaust gas
measurements. Please take the matter up with anapol.
- IMPORTANT NOTE -
If your device is equipped with an automatic CO cut-off system, you can still
continue measuring, because the CO sensor will automatically be flushed out
with fresh air.
9.5
Error message during printing or when saving data to a chip card
No data for chip
card
E : Repeat
Cause: there is no data in the device's memory.
Solution: in the "Settings" menu -
,
, the "Data
logging" function must be set to "Yes".
In the "Communication" menu -
,
,
, the
"Data output" function must not be selected.
Chip card
not ready
E : Repeat
P : Cancel
Cause: the chip card has not been recognised.
Solution: check that the chip card has been inserted
correctly. Clean the contacts. Try again with a different
card.
9.6
Error message during soot measurement
Error
Soot reading > 3
E : Repeat
Cause: there is too much gas flow resistance.
Solution: check if the soot filter is wet, there is too much
soot or the probe is blocked.