APPH6000 User Manual V1.0 2010
- | Introduction to the APPH6000
7
Front Panel Overview
Depending on the APPH6000 product configuration, the front panel can contain up to four
female SMA connectors and five female BNC connectors. A green LED indicates the power
on/off status of the instrument. Figure 1 shows the front panel of the APPH6000 with option
CC. For a single channel model, the second channel IOs are sealed.
RF IN(1) connector
This female SMA connector is the DUT signal input. The impedance is 50
ohm. The damage level is +36 dBm maximum. The maximum allowed DC level is +/- 10 V.
RF IN2 connector (optional)
This female SMA connector is the second DUT signal input in case
the two-channel cross-correlation option is installed. When using the cross-correlator, the DUT
signal is split to provide equal amplitudes at RF1 and RF2 inputs. The impedance is 50 ohm. The
damage level is +36 dBm maximum. The maximum allowed DC level is +/- 10 V.
REF IN(1) connector
This female SMA connector is the reference LO input. The impedance is 50
ohm. The damage level is +36 dBm maximum. The maximum allowed DC level is +/- 10 V.
Figure 1 APPH6000 Front Panel View (APPH6000 instrument with option CC)