Care, maintenance, lamp change
Changing the desiccator
8.4
Changing the desiccator
Danger! Hot surface!
Before replacing the desiccator, switch off the SPECORD
®
. Allow for a sufficient cool-
down time of the lamps. Touching the hot lamp surfaces housings may cause burns!
The desiccator protects the optical system of the SPECORD
®
against too high air
humidity. The desiccator is accessible, when the lamp compartment has been opened
(
→
Especially in tropical countries, the adsorption capacity of the desiccator should be
checked regularly. The adsorption capacity of the desiccator is exhausted, if the color of
the indicator paper in the inspection window has changed from blue to pink-yellow.
For changing the desiccator, follow this procedure:
1. Open the flap of the lamp compartment by folding the lock frontward so it sticks out
vertically. Lift the flap from below by 1 to 2 mm and then pull it off to the left.
2. Using a flat tool, such as an appropriate coin, unscrew the desiccator from its
holder.
3. Remove the new desiccator from the packaging and screw it into the holder.
4. Finally, close the flap of the lamp compartment.
Used up desiccators can be reused after heating them up for five hours in a drying oven.
New desiccators are available from our Service Department (
→
for the address, refer to
the inside cover page).
8.5 Changing
fuses
If the line fuses are defective, you can change them yourself.
Risk of electric shock!
Before changing the fuses, make sure to switch the SPECORD
®
off with the power
switch and disconnect the power cable from the power input connector of the device.
1. Remove the power cable from the SPECORD
®
.
2. Open the fuse holder (
→
Refer to Fig. 6-3, p. 22) by pulling on its lid.
3. Replace the defective line fuses.
Make sure to use only the following fuse types:
•
T 3.15 H/250 V, Type 19181 for line voltages of 150 VAC and 230 VAC.
4. Close the fuse holder again.
5. Reconnect the power cable to the power input connector of the SPECORD
®
and the
power outlet.
6. Switch the SPECORD
®
on again.