Care, maintenance, lamp change
Changing lamps
8.3.4 Changing the halogen lamp
Electric shock!
Before changing the halogen lamp, make sure to switch the device off with the power
switch.
Caution: Risk of burns!
Allow the halogen lamp to cool down to avoid the risk of burns!
Emitted radiation!
Never look into the lighting lamp as the radiation may cause eye injury!
Caution!
Do not touch the glass bulb of the lamp to avoid any contamination affecting its
performance!
Open the lamp-compartment flap. The protective cover of the halogen lamp is visible.
1. Loosen the knurled screw of the protective cover of the halogen lamp (Fig. 8-2, 2)
and remove the front part of the protective cover.
Note
The protective cover of the halogen lamp of the SPECORD® 200/205/210 was
technically revised. On devices produced before, the protective cover consists only of a
single part. In this case, remove the complete protective cover upward.
1
Rear part of protective cover of halogen lamp
2
Knurled screws
4
Holder of halogen lamp
3
Adjusting screws for halogen lamp holder
4 Halogen
lamp
Fig. 8-1
Changing the halogen lamp