Installation and start-up
Start-up
Note
When required, align the lamps following the instructions given in Section 8.3 Changing
lamps, p. 32 after start-up.
6.4 Start-up
Do not cover ventilation slots of the device!
Take care to ensure that the ventilation slots on the top and on the right and left
bottom of the SPECORD
®
are not covered. Do not place any objects in front of the
openings. The suction intakes for the two fans on the bottom of the device must be
kept free.
1. Start
WinASPECT
®
software by a click on the program icon on the Windows
desktop or by clicking the Start button on the taskbar:
Start / Programs / WinASPECT / WinASPECT
.
Fig. 6-7
Program symbol of WinASPECT
®
2. Using the power switch on the left panel of the device, switch the SPECORD® on
(
→
Refer to Fig. 6-3 Connectors on left-hand side panel, p. 22).
3. The green Power ON LED on the front panel of the device lights.
(
→
Refer to Fig. 6-2 Display elements on front panel, p. 21).
4. In WinASPECT® software, select the COM port for the SPECORD® – PC
connection:
•
Activate the menu command
Measurement / Set Parameters
to open the
window of the device driver
SPECORD xxx ....
Fig. 6-8
Dialog box of SPECORD
®
250 device driver
•
Click on the
Device
tab.
SPECORD
®
200/205/210/250 User's Manual – Edition 04/2005
Page 25