Installation and commissioning
25
4.3
Installing and connecting the device
WARNING
Risk of electric shock! Do not touch!
NEVER connect or separate the power cable from the power connector when the green
status LED is lighting on the power module and the qTOWER
3
auto. Pay attention to the
warning notes in the section “Safety instructions: during operation” on page 11.
Place the qTOWER³ auto and the power module on a sturdy, even surface in a secure,
dry place.
Leave the device switched off to allow it to acclimatize adequately (> 2 hours).
1.
Make sure the power switch at the back of the power module is OFF before connect-
ing the power cable to a grounded power outlet (100 V to 230 V). Take note of the
voltage rating on the type plate beside the power cable on the back of the power
module.
Note:
Make sure that the power switch and the power cable are easily accessible.
This is important in the event that the device needs to be separated from the supply
voltage.
Fig. 11 Connecting the qTOWER
3
auto