16
AC & AC+DC Voltage
RANGE AM-160 AM-140
Accuracy*
20Hz -- 45Hz
500.00mV 1.5% + 40d
5.0000V,
50.000V
1.5% + 40d
500.00V,
1000.0V
Unspec'd
Unspec'd
45Hz -- 300Hz
500.00mV 0.3% + 20d
5.0000V,
50.000V
0.8% + 20d
500.00V,
1000.0V
0.4% + 40d
0.8%+60d
300Hz -- 5kHz 300Hz -- 1kHz
500.00mV 0.3% + 10d
0.8%+40d
5.0000V,
50.000V,
500.00V
0.4% + 40d
2.0%+60d
1000.0V
0.8% + 40d
(300Hz--1kHz)
1.0%+40d
5kHz -- 20kHz 1kHz -- 20kHz
500.00mV 0.5%+20d
1dB**
5.0000V,
50.000V
0.8%+20d 2dB**
500.00V 0.5%+20d 3dB**
1000.0V Unspec'd Unspec'd
20kHz -- 100kHz
500.00mV 2.0%+40d
5.0000V,
50.000V
4.0%+40d**
500.00V
1000.0V
Unspec'd
Unspec'd
*From 5% to 10% of range:
accuracy % of reading (or in dB) + 80d
**From 5% to 10% of range:
accuracy % of reading (or in dB) + 180d
**From 10% to 15% of range:
accuracy % of reading (or in dB) + 100d
CMRR: >90dB @ DC to 60Hz, Rs=1k
Ω
Input Impedance: 10M
Ω
, 30pF nominal
(80pF nominal for 500mV range)
Residual reading less than 50 digits with
test leads shorted.
dBM
At 600
Ω
, -11.76dBm to 54.25dBm,
Accuracy:
±
0.25dB + 2d (@40Hz --
20kHz)
Input Impedance: 10M
Ω
, 30pF nominal
Selectable reference impedance of 4, 8,
16, 32, 50, 75, 93, 110, 125, 135, 150,
200, 250, 300, 500, 600, 800, 900, 1000,
1200
Ω
DC Voltage
RANGE AM-160
AM-140
Accuracy
500.00 mV,
5.0000V,
50.000V
0.02% + 2d 0.03% + 2d
500.00V
0.04%+2d 0.05% + 2d
1000.0V 0.05%+
2d
0.1%+2d
NMRR: >60dB @ 50/60Hz
CMRR: >120dB @ DC, 50/60Hz, Rs=1k
Ω
Input Impedance: 10M
Ω
, 30pF
nominal (80pF nominal for 500mV
range)
Audible Continuity Tester
Audible threshold: between 20
Ω
and
200
Ω
Response time < 100
µ
s