8 Controller Operation
35
8
Controller Operation
This chapter discusses in detail how to operate the bath temperature controller
using the front control panel. Using the front panel key switches and LED dis-
play the user may monitor the bath temperature, set the temperature set-point in
degrees C or F, monitor the heater output power, adjust the controller propor-
tional band, set the cutout set-point, and program the probe calibration parame-
ters, operating parameters, serial interface configuration, and controller
calibration parameters. Operation is summarized in Figure 5 on page 36.
8.1
Bath Temperature
The digital LED display on the front panel allows direct viewing of the actual
bath temperature. This temperature value is what is normally shown on the dis-
play. The units, C or F, of the temperature value are displayed at the right. For
example,
Bath temperature in degrees Celsius
The temperature display function may be accessed from any other function by
pressing the “EXIT” button.
8.2
Reset Cutout
If the over-temperature cutout has been triggered then the temperature display
will alternately flash,
Indicates cutout condition
The message continues to flash between the actual temperature and
Cut-out
until the temperature is reduced and the cutout is reset.
The cutout has two modes — automatic reset and manual reset. The mode de-
termines how the cutout is reset which allows the bath to heat up again. When
in automatic mode, the cutout will reset itself as soon as the temperature is low-
ered below the cutout set-point. With manual reset mode the cutout must be re-
set by the operator after the temperature falls below the set-point.
When the cutout is active and the cutout mode is set to manual (“reset”) then
the display will flash “cutout” until the user resets the cutout. To access the re-
set cutout function press the “SET” button.
S
Access cutout reset function
The display indicates the reset function.
25.00 C
Cut-out
Summary of Contents for Kaye CTR-40
Page 1: ...M4436 Rev B 12 19 Kaye CTR 40 Cold Temperature Reference User s Guide...
Page 2: ......
Page 11: ......
Page 17: ...1 Before You Start 6 Voltage Cut In 7 5 213 247 VAC...
Page 19: ......
Page 27: ......
Page 32: ......
Page 45: ......
Page 65: ......
Page 73: ......
Page 83: ...72 12 Troubleshooting 12 3 Wiring Diagram Figure 9 Wiring Diagram...