2 Introduction
9
3
Specifications and Environmental Conditions
3.1
Specifications
–40°C to 150°C
Stability (2 sigma)
±0.005°C at –40°C (ethanol)
±0.005°C at 25°C (water)
±0.007°C at 150°C (5012 oil)
Uniformity
±0.006°C at –40°C (ethanol)
±0.005°C at 25°C (water)
±0.010°C at 150°C (5012 oil)
Heating Time†
60 minutes, from 25°C to 150°C (5012 oil @ 115V)
Cooling Time
110 minutes, from 25°C to –40°C (ethanol)
Stabilization Time
15-20 minutes
Temperature Setting
Digital display with push-button entry
Set-point Resolution
0.01°; 0.00018° in high resolution
Display Temperature Resolution
0.01°
Digital Setting Accuracy
±0.5°C
Digital Setting Repeatability
±0.01°C
Heater
1400 VA
@ 115V (230V) Nominal
Access Opening
172mm x 94mm (6.8” x 3.7”)
Depth
234 mm (9.25”)
Wetted Parts
304 Stainless Steel
Power †
115 V ac (±10 %), 60 Hz, 15 A [230 V ac (±10 %), 50 Hz, 8 A optional], 1400 VA
Caution
W
If the voltage is outside ±10%, the compressor may be damaged.
Check the
back panel label for the correct voltage and frequency prior to energizing
the unit.
Volume
9.2 liters (2.4 gal.)
Weight
35.4 kg (78 lb.)
Size
305 mm
W
x 622 mm
D
x 584 mm
H
(12” x 24.5” x 23”) (W x D x H)
Safety
Refrigeration
R-
410A
single stage
Interface Package
RS-232 included
†Rated at listed 115 V (or optional 230 V).
IEC 61010-1; Overvoltage Category II, Pollution Degree 2; IEC 61010-2-010
,
IEC 61010-2-011
Electromagnetic Compatibility
International ........................................IEC 61326-1: Basic Electromagnetic
Environment; CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-cou-
pled radio frequency energy that is necessary for the internal function of the
equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic
and those directly connected to a low-voltage power supply network that supplies
buildings used for domestic purposes. There may be potential difficulties in
ensuring electromagnetic compatibility in other environments due to conducted
and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and
may not provide adequate protection to radio reception in such environments.
Range
Summary of Contents for Kaye CTR-40
Page 1: ...M4436 Rev B 12 19 Kaye CTR 40 Cold Temperature Reference User s Guide...
Page 2: ......
Page 11: ......
Page 17: ...1 Before You Start 6 Voltage Cut In 7 5 213 247 VAC...
Page 19: ......
Page 27: ......
Page 32: ......
Page 45: ......
Page 65: ......
Page 73: ......
Page 83: ...72 12 Troubleshooting 12 3 Wiring Diagram Figure 9 Wiring Diagram...