Maintenance and Troubleshooting
|
5-7
bench testing
When the fault has been resolved, the probe should be bench-tested before fitting
it back into the flue.
1. Allow the probe to operate in air for about 20 minutes in order to achieve tem-
perature stability.
2. Periodically check the cell EMF reading after the initial stabilizing time has
expired.
3. The reading should settle at the cell constant of typically 0V ±2mV for a new cell.
4. When the cell attains its constant (typically 1 hour), apply a zero calibration gas
and check the cell EMF. If necessary, carry out a full calibration.
The cell can take up to three hours to fully settle.
If the cell responds correctly to the calibration gas, the probe can be refitted into
the flue.