4-22 | Thermox WDG-1200/1210 Insitu Probes
register numbers
This manual uses Register Numbers as defined in the Modbus specifications.
The first register number is 1. Modbus packets contain Register offsets which
are one less than register Numbers and start at offset 0.
data types
Integer variable
Single register.
registers numbers for Instrument readings (integer value)
address
Variable
scaling
23
O
2
Reading
1000
27
Cell Temperature (
o
C)
10
command register
address
Variable
scaling
30
Status and Calibration Start
1
To start the calibration sequence, write the value 1 to register 30.
Maintenance status register
address
Variable
scaling
19
Maintenance State
1
The value of Register 19 is 1 when the Maintenance Relay is on; otherwise it is
0.