User Interface / Calibration
|
4-21
Modbus Interface
serial data bus configuration
The serial data bus can be set to any of the following four configurations:
code
baud rate
Parity
stop bits
1200
1200
Even
1
4800
4800
Even
1
9600
9600
Even
1
96nP
9600
None
1
Data format: Modbus RTU
The default setting is 1200. To chnage the serial data setting:
Press F1. Use the up arrow until the display reads 417 and press Enter.
Press Enter again and F1 four times until the display shows: SYSt.
Press Enter and F1 until the display shows: bAud.
Press Enter and use the up and down arrows to change the baud code.
When correct, press and hold Enter until the display flashes.
Proceed to next section.
slave address and floating Point format
The default slave address is 11. To change the slave address, first set the baud
rate as in previous section, then:
Press F1. The display shows: PrAd.
Press enter and use the up and down arrows to change the slave address.
When correct, press and hold Enter until the display flashes.
Press F1 twice to return to the normal display.
function codes supported
03
Read multiple register
06
Pre-set single register
exception codes
01
Illegal function
02
Illegal address
03
Illegal data value (Write request to a read-only customer)
Summary of Contents for Thermox WDG-1200 Insitu
Page 9: ...ix...
Page 15: ...1 6 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Page 19: ...2 4 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...
Page 86: ...Maintenance and Troubleshooting 5 13 This page intentionally left blank...
Page 98: ...6 12 Thermox WDG 1200 1210 Insitu Probes This page intentionally left blank...