Surtronic S25
is located in the front panel of the Display Unit, just
above the pick-up connector.
5.
Repeat the measurement and adjustment, until the measured
value is within 2% of the value marked on the specimen.
Cleaning the Stylus
Occasionally clean the stylus with a camel hair brush moistened
with a proprietary cleaning agent.
Pick-up Skid
To reduce the effect of wear, on pick-ups with a rotatable skid,
occasionally turn the skid round to present a new contact surface.
The skid is clamped by the screw in the front of the pick-up.
When loosening this screw, take care not to let the retaining pin
next to the skid fall out. When retightening the screw, ensure that
the end cover is correctly positioned with respect o the stylus and
that the stylus is free to move.