Surtronic S25
Operating Error Indications
During a measurement a message may be displayed which
indicates that an error condition has occurred. The messages and
the probable causes for their display are as follows:
Display
Reason for message
Motor Error Motor fault
Pick-up Error Response at pick-up level error or faulty
connection
Over range
Selected range too small
E
Can occur in front of a measured value due to
excessive slew rate for a single data point (large
spike). May also occur if the pick-up is lifted off
the surface
Data transmit
error Printer
connected
at
start of transmission but
later detected as not connected (printer off line)
Battery low When battery voltage <6.4V: key push
operations are disabled and display is off.
When the instrument is in the sleep mode and
the Measure key is pressed, then "Battery low" is
displayed for 2 seconds, followed by the normal
display or operation.
Stored surface profile is erased
Renew or recharge battery
Page 5-4
Summary of Contents for Surtronic 25
Page 2: ......
Page 8: ...Surtronic S25 Page vi...
Page 14: ...Surtronic S25 Page 1 6...
Page 24: ...Surtronic S25 Page 2 10...
Page 30: ...Surtronic S25 Page 3 6...
Page 38: ...Surtronic S25 Page 4 8...
Page 48: ...Surtronic S25 Page 5 10...
Page 58: ...Surtronic S25 Page 6 10...