Transportation/Setting-up
— 12
SPECTRO MIDEX — 31.08.2016
4 Transportation/Setting-up
The instrument - in its transport packaging - may only be transported by
specialized organizations or by specifically trained personnel.
Unpacking, installation, setup and connection of the instrument may only be
performed by service personnel that is authorized by the manufacturer.
Note! Measurement errors!
Shocks to the instrument during the measuring process may result in
measurement errors.
Therefore, shocks to the instrument when taking measurements must
be avoided.
Avoid mechanical impacts.
Measurement errors may result from temperature differences inside the
instrument.
Install the instrument so that it cannot heat up on one side only.
Observe the following information regarding the instrument's installation/transport:
Always keep the instrument upright when transporting it.
Always use the handles to carry the instrument (see also page 6/Fig.
1/point 6).
Save the shipping box in case you have to ship the instrument to the service
department at a later date.
When setting up the spectrometer, ensure that all sides are easy to access for
maintenance, repair and connection/disconnection purposes.
Set up the instrument to ensure the power isolators (switches and power
cable) can be easily reached.
All necessary cables and hoses are included in the delivery.
If the instrument is to be moved to another place of installation after set-up,
there is a risk of damaging the instrument. Contact the SPECTRO Service
Department in this matter.