Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-2
M370185-01
Introduction
The rear panel connectors J1 and J3 provide an option to control and
monitor the output of the power supply with analog signals. Connector J1
provides a non-isolated analog interface where all signals are referenced
to the negative output terminal of the power supply. Connector J3 is an
isolated interface and also provides an isolated auxiliary voltage output to
aid analog programming.
Analog Programming (APG) of Output Voltage and Output Current
Analog programming allows control of the power supply’s output
voltage, output current or both to be controlled from analog input sources.
As the programming source is varied within the configured analog
programming range (2–10 volts/
k
Ω
max), the power supply’s output
varies proportionally over its output range. For example, if you set the
programming scale to 10 V voltage source and apply 5 V to the
programming pins, the power supply will be programmed to 50% of the
output capability for your power supply.
While analog programming mode is configured, you will not be able to
make adjustments to the set points from the front panel or through the
remote interfaces.
The front panel will display
AnPr
in the output current display when the
9-position mode control is turned to VOLTS or AMPS positions for the
parameter(s) that are configured to be controlled by analog programming.
If an adjustment of a parameter controlled by APG is attempted using a
SCPI command, a “-221 Settings Conflict” error will be pushed onto the
error queue.
Using analog programming requires that you make connections to the
Isolated Analog Programming Connector or Analog Programming
Connector on the rear panel of the power supply. Depending on your
model, see “Rear Panel Connectors” on page 1–6 or “” on page 1–7.
Summary of Contents for Soresen XTR 100-8.5
Page 2: ......
Page 16: ...xiv ...
Page 28: ...1 8 ...
Page 50: ...Local Operation 3 8 M370185 01 Figure 3 3 Front Panel Menu System ...
Page 93: ...Using Multiple Power Supplies M370185 01 3 51 3 ...
Page 94: ...Local Operation 3 52 M370185 01 ...
Page 128: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370185 01 ...
Page 198: ...Remote Operation 5 70 M370185 01 ...
Page 261: ...SCPI Command Tree M370185 01 A 35 A ...
Page 262: ...A 36 ...
Page 279: ...Index M370185 01 IX 3 described 3 7 3 36 number available 3 36 stored values 3 36 ...
Page 280: ...IX 4 ...