
24
Installation, Operation and Maintenance Manual
OXYvisor Optical Oxygen Analyzer
The OXYvisor Optical Oxygen Analyzer includes a 316 SS pipe-mount bracket, or field mounting. An optional
wall mount kit is available. Additional components needed to produce a functional oxygen measurement
include a Barben Oxygen Sensor (BOS) and Pt1000 RTD (Resistance Temperature Sensor). Other
components are commonly required to make a reliable measurement; such as a pressure transmitter, “wet-
gas” sample probe, heated pressure reduction stations (HPRS), heated sample lines (HSL’s) and a manual
or automatic sample calibration panel (SCP). If any of these items are required, please consult a Barben
Analytical representative regarding your application.
It is important that the product
Name Plate
(Figure 6) is reviewed to properly identify the product, it’s power input,
and the specific hazardous area markings for the actual unit. It is equally important that the installation control
drawing referenced in Table 5, is fully understood and followed along with the inclusion of any applicable
federal, state and local codes.
Section 4 - Installation
4.1 Installation Overview
4.2 Inspection of the Analyzer Shipment (consignment)
Check the packaging and the device for visible damage caused by inappropriate handling during shipping.
Close attention should be paid to the powder coated analyzer enclosure for signs of impact or abrasion during
shipping. All threads should be inspected for burrs and any residual packing material should be removed.
Report any claims for damages immediately to the shipping company. Take pictures and retain damaged
parts for clarification. Check the scope of delivery by comparing the shipping documents with your order for
correctness and completeness.
The OXYvisor is used to quantitatively measure the amount of oxygen in a gaseous or liquid sample. The
analyzer may only be installed as specified in these instructions per its intended use. Modifications to the
analyzer which are not expressly referred to within these instructions will result in an application which is not in
accordance with the intended use. Such modifications are then the exclusive responsibility of the user.
Do not use any damaged or incomplete devices. Danger of explosion in hazardous areas.
OXYvisor Model Number Type of Protection Installation - Control DWG
BOA-DC-1-SFP-B-SI
Zone 1 IIC
2P0345 (Zone 1)
BOA-AC-1-SFP-B-SI
BOA-DC-1-SFP-B-AM
BOA-AC-1-SFP-B-AM
BOA-DC-2-SFP-B-SI
Zone 2 IIC
&
Class I, Division 2
Groups A,B,C,D
2P0346 (Zone 2)
2P0335 (C lass I Div2)
BOA-AC-2-SFP-B-SI
BOA-DC-2-SFP-B-AM
BOA-AC-2-SFP-B-AM
!
WARNING
Table 5 - OXYvisor HazLoc Summary by Product Model Number Selection
Section 4
Installation