Specifications
AQUATRAN 3/38 Operator’s Manual
9-4
Revision G
MOCON, Inc.
Transmission Rate Range and Repeatability Capabilities
Testing Range
Carrier
Flow
Minimum
Exam
Time
Required
Calibration
Standard
Repeatability
2.5 to 100 gm/(m
2
-day)
0.161 to 6.45 gm/(100
2
-day)
0.0125 to 0.50 gm/(pkg-day)
100
sccm
25
minutes
Certified Film
P/N 054-162
(yellow)
+/- 2.0% of Reading
0.05 to 2.5 gm/(m
2
-day)
0.0032 to 0.161 gm/(100
2
-day)
0.00025 to 0.0125 gm/(pkg-day)
100
sccm
25
minutes
Certified Film
P/N 054-163
(Red)
+/-0.05 g/(m
2
• day)
0.05 to 10 gm/(m
2
-day)
0.0032 to 0.645 gm/(100
2
-day)
0.00025 to 0.05 gm/(pkg-day)
10
sccm
45
minutes
Certified Film
P/N 054-164
(
Blue
)
+/- 2.0% of Reading or
+/-0.05 g/(m
2
• day)
whichever greater
Table 9-6: Range and Repeatability Capability
Notes:
1.
The ReZero Frequency must be set to one.
2.
When testing samples in both cells the transmission rates must be within a factor of 5.
3.
Calibration will be performed at the flow rate the samples were tested at.
4.
Testing performed at 37.8 Degrees Centigrade.
5.
The Repeatability specifications assume the use of a 100% RH Test Gas.
6.
Repeatability is based on a film area of 50 cm
2
.