Safety Features
6
LD Series Operation Manual
CROSSHEAD MOVEMENT LIMITS
The LD Series is fitted with two mechanically activated limit stops. These can be used as extra protection to stop load cells, grips or fixtures coming into con-
tact, or to back up the tests upper and lower limits. Activating a limit stop will automatically stop the machine.
X
To Adjust a Limit Stop
1
Loosen the limit stop screw, move it to the top or bottom of its travel, and tighten the screw temporarily.
2
Drive the moving crosshead to the highest or lowest position it may go to safely, without damaging any fitted grips or fixtures.
3
Move the crosshead slightly away from the position established in step 2.
4
Loosen the screw and move it to the cross head actuator and tighten in this position.
Move the Limit Stop to the top or bottom
drive the Moving Crosshead to a position
then move the Limit Stop to the Crosshead
of its travel…
it may go to safely…
Actuator position and tighten it.
OPTIONAL SPLINTER SHIELD
A transparent splinter shield is available as an option to enclose the test sample. A shield should be
used if any danger of samples shattering during the test exists. The shield has an electronic interlock,
which requires the hinged shield to be closed before the test can start.
SAFETY JOG
The LD Series has a built in safety feature to provide protection while jogging the machine. If the
machine measures an unexpected resistance while jogging, it will automatically stop.
Note:
The resistance level is selectable in the Nexygen
Plus
4 software.
Optional Splinter Shield.