Safety Features
11
LD Series Operation Manual
MAINS INPUT VOLTAGE SELECTION
The LD Series can be used with mains power of either 230VAC ±10% or 115VAC ±10%, with frequency from 50Hz to 60Hz.
!
CAUTION:
Check to ensure the correct voltage of your mains power supply has been selected on the input connector BEFORE connecting power to the machine.
X
To Set the Correct Mains Power Supply Voltage
1
Remove the rear plastic housing shroud from the machine.
2
Remove the protective cover over the terminal strip, and identify the four terminals on the left end of the terminal strip located on the back of the machine.
Rubber mat removed for clar
ity.
+2
opposite
side
Location of Terminal Strip
Remove the 13 screws (
) to remove the rear plastic housing shroud…
then remove the terminal strip protective cover.
3
To configure for 115 volt mains, configure the two jumpers between the outer screw terminals.
4
To configure for 230 volt mains, configure a single jumper between the inner screw terminals.
Configuration for 115 volt mains.
Configuration for 230 volt mains.
5
Replace the protective cover over the terminal strip.
6
Replace the rear plastic housing shroud.