![Ametek LD10 Operation Manual Download Page 15](http://html1.mh-extra.com/html/ametek/ld10/ld10_operation-manual_2931370015.webp)
Safety Features
13
LD Series Operation Manual
LOAD CELLS AND GRIPS
X
To Install a Load Cell
1
Select the correct size load cell adapter.
2
Place the adapter through the hole in the moving crosshead with the thread side down.
3
Screw the load cell adapter into the load cell.
Note:
Take care to ensure that the locating pin on the load cell fits in one of the holes on the underside of the moving crosshead.
Place the adapter into the moving crosshead… then screw the adapter into the load cell.
4
Check that the connecting cable is on the same side as the connecting socket, then tighten the load cell adapter with the C spanner provided.
5
Insert the load cell plug in the socket on the rear of the moving crosshead, and finger-tighten the two retaining screws.
6
Fit the eye end to the load cell and tighten with the C spanner.
Use
C-spanner
wrench
here