3.1 CJC Control
To disable the CJC proceed with the following steps:
1) With the unit turned off press and hold the I/O key.
2) Turn on the unit while still holding down the I/O key.
3) Approximately 5 seconds after power up release the I/O key. The CJC
off annunciator will be displayed on the LCD when operated in the TC
mode.
4) Turning the CSC200 off then back on will return the unit to normal
operation (CJC on).
NOTE:
When operating in the direct millivolt mode the CJC is not used.
Figure 2
4. Operation in the RTD (ohms) Mode
The CSC200 can input or output any one of 17 different RTD types including
ohms. Operation in the input mode includes 2, 3 and 4 wire RTD's and YSI
Series 400 Thermistors. When operating in the output mode the CSC200
can simulate a RTD output into virtually all transmitters and RTD input
devices including pulsed excitation current instruments.
4.1
RTD Input (measure) Mode Operation
To operate the CSC200 in the input mode refer to Figure 3 for the proper
input connections and proceed as follows:
1) Select the RTD mode by pressing the "TC/RTD" key.
2) Select the RTD type by holding the "Type" key and turning the knob.
Note:
The PT-100 (platinum 100 ohm) has several Alpha curve choices
(385, 392, JIS, etc.), take care to select the proper one for your
application.
Note:
The TC wire used must
match the thermocouple type
being calibrated.