13
Thermocouple Read and Source (errors in °C / °F)
TC Type
Range (°C / °F)
Accuracy (°C / °F)
J
-210.0 - 0.0 / -346 - 32
0.6 / 1.08
0.0 - 800.0 / 32 - 1472
0.4 / 0.72
800.0 - 1200.0 / 1472 - 2192
0.5 / 0.90
K
-200.0 - 0.0 / -346 - 32
0.8 / 1.44
0.0 - 1000.0 / 32 - 1832
0.5 / 0.90
1000.0 - 1372.0 / 1832 - 2501.6
0.7 / 1.26
T
-250.0 - 0.0 / -418 - 32
0.8 / 1.44
0.0 - 400.0 / 32 - 752
0.4 / 0.72
E
-250.0 - -100.0 / -418 - -148
0.8 / 1.44
-100.0 - 1000.0 / -148 - 1832
0.4 / 0.72
R
0.0 - 1767.0 / 32 - 3212.6
1.4 / 2.52
S
0.0 - 1767.0 / 32 - 3212.6
1.4 / 2.52
B
600.0 - 800.0 / 1112 / 1472
1.4 / 2.52
800.0 - 1000.0 / 1112 - 1832
1.5 / 2.70
1000.0 - 1820.0 / 1832 - 3308
1.7 / 3.06
C
0.0 - 1000.0 / 32 - 1832
0.8 / 1.44
1000.0 - 2316.0 / 1832 - 4200.8
2.5 / 4.50
XK
-200.0 - 800.0 / -328 - 1472
0.4 / 0.72
BP
0.0 - 800.0 / 32 - 1472
1.1 / 1.98
800.0 - 2500.0 / 1472 - 4532
2.5 / 4.50
L
-200.0 - 0.0 / -328 - 32
0.45 / 0.81
0.0 - 900.0 / 32 - 1652
0.4 / 0.72
U
-200.0 - 0.0 / -328 - 32
0.7 / 1.26
0.0 - 600.0 / 32 - 1112
0.45 / 0.81
N
-200.0 - 0.0 / -328 - 32
1.0 / 1.80
0.0 - 1300.0 / 32 - 2372
0.6 / 1.08
All TC errors include CJC errors
CJC error outside of 23 ± 5°C is 0.05°C/°C (73.4 ± 9°F is 0.09°F/°F)
Note: All accuracies stated are for operation at 23°C ±5°C (73.4 ± 9°F).
For operation below 18°C or above 28°C add ±0.005% of reading/°C (below 64.4°F
or above 82.4°F add ±0.003%/°F).