Ametek JOFRA CSC200 Reference Manual Download Page 6

4

Caution

To avoid possible damage to calibrator or to equipment under test:

• Disconnect the power and discharge all high-voltage capacitors before

testing resistance or continuity.

• Use the proper jacks, function, and range for your measurement or

sourcing application.

• To avoid damaging the plastic lens and case, do not use solvents or

abrasive cleansers.

Clean the calibrator with a soft cloth dampened with water or water and mild
soap.

2. Calibrator Interface

2.1 Display and Inputs

The CSC200 has a large LCD area that displays the function selected and
the input/output value.  Input terminals on the top of the calibrator are used
for all TC and RTD connections Figure 1 illustrates many if these features.

2.2 Key Functions

Figure 1 shows the layout of the calibrator keypad while Table 1 lists the
calibrator keys and their functions.

Summary of Contents for JOFRA CSC200

Page 1: ...ReferenceManual CompactSignalCalibrator AMETEKJOFRACSC200 Copyright 2005AMETEK DenmarkA S...

Page 2: ......

Page 3: ...ll Setpoints are stored in non volatile memory and will remain until new setpoints are stored 1 1 Contacting Ametek US Canada Latin America AMETEK TCI at 1 800 527 9999 Europe Africa Middle East AMETE...

Page 4: ...se symbols may be used on the instrument or in this manual Symbol Description AC Alternating Current AC DC Battery CE Complies with European Union Directives DC Double Insulated Electric Shock Fuse PE...

Page 5: ...n and range for your measurement Make sure the battery cover is closed and latched before you operate the calibrator Remove test leads from the calibrator before you open the battery door Inspect the...

Page 6: ...use solvents or abrasive cleansers Clean the calibrator with a soft cloth dampened with water or water and mild soap 2 Calibrator Interface 2 1 Display and Inputs The CSC200 has a large LCD area that...

Page 7: ...r up function 3 Displays and selects TC or RTD type used in conjunction with knob 4 Selects input measure or output source mode 5 Recalls stored setpoints for the function selected 6 Press to store se...

Page 8: ...latile memory until changed by the user The unit ships from the factory with the temperature units set in the C mode To change to F do the following steps 1 With the CSC200 turned off hold down the TC...

Page 9: ...erent RTD types including ohms Operation in the input mode includes 2 3 and 4 wire RTD s and YSI Series 400 Thermistors When operating in the output mode the CSC200 can simulate a RTD output into virt...

Page 10: ...SC200 in the output mode refer to Figure 4 for proper output connections and proceed as follows 1 Select the RTD mode by pressing the TC RTD key 2 Select the RTD type by holding the Type key and turni...

Page 11: ...and 17 RTD functions Stored setpoints will remain in non volatile memory until changed by the operator To store and recall setpoints proceed as follows 5 1 Storing Setpoints 1 Select an output functio...

Page 12: ...f 23 5 C Power 1 x 9V Alkaline Low Battery Indicator 7V approximate Size Approx 188mm x 84mm x 52mm 7 37in x 3 31 in x 2 05in Weight Approx 400g 14oz Note All accuracies stated are for operation at 23...

Page 13: ...385 200 0 100 0 328 212 0 8 1 44 100 0 300 0 212 572 0 9 1 62 300 0 630 0 572 1166 1 0 1 80 P400 90 385 200 0 100 0 328 212 0 4 0 72 100 0 300 0 212 572 0 5 0 90 300 0 630 0 572 1166 0 6 1 08 P500 90...

Page 14: ...212 0 2 0 36 100 0 300 0 212 572 0 3 0 54 300 0 630 0 572 1166 0 4 0 72 M10 90 427 100 0 260 0 148 500 1 4 2 52 M50 90 428 180 0 200 0 292 392 0 4 0 72 M100 90 428 180 0 200 0 292 392 0 3 0 54 H120 9...

Page 15: ...2 52 800 0 1000 0 1112 1832 1 5 2 70 1000 0 1820 0 1832 3308 1 7 3 06 C 0 0 1000 0 32 1832 0 8 1 44 1000 0 2316 0 1832 4200 8 2 5 4 50 XK 200 0 800 0 328 1472 0 4 0 72 BP 0 0 800 0 32 1472 1 1 1 98 8...

Page 16: ...solvents or abrasive cleansers Clean the calibrator with a soft cloth dampened with water or water and mild soap 7 3 Service Center Calibration or Repair Only qualified service personnel should perfor...

Page 17: ...re adapter kit Type T thermocouple 121985 Extension Cable for Pt 100 Sensor 5 m 121983 Extension Cable for Type K 5 m 122523 Extension Cable for Type N 5 m 120519 Thermocouple Male Plug Type Cu Cu Whi...

Page 18: ......

Page 19: ...SPK SCS 002 7 05 Rev C 0219130...

Page 20: ...nge without notice All rights reserved because calibration is a matter of confidence AMETEK Calibration Instruments offers a complete range of calibration equipment for pressure temperature and signal...

Reviews: