Figure 4-4
Testing Integral Cable
4.7
Testing the Integral Cable
4.8
Testing the Remote Cable
1. Disconnect integral cable from electronic.
2. Check for continuity.
a. Using an ohmmeter measure the resistances.
i. From the center wire connection on the RCA plug to
the sensing element active section (center wire)
ii. From the shield connection on the RCA plug to the
sensing element shield.
iii. If resistance is greater than 5Ω the cable has failed.
3. Check for shorts.
a. The integral cable is tested as part of sensing element
test,
section 4.6
.
1. Disconnect remote cable from electronic unit and sensing
element.
2. Using an ohmmeter measure the resistances as
shown in
Figure 4.5
a. Check for shorts.
i. Connect ohmmeter to cable center wire and ground
wire. Measure resistance as shown.
ii. Move ohmmeter leads and repeat for all measurements
shown.
iii. If resistance is less than 100 k-ohms cable has failed
b. Check for continuity.
i. Short center wire to ground wire and measure
resistance as shown. Should be close to 0 ohms
ii. Short center wire to shield and repeat.
iii. If resistance is greater than 10 ohms cable has failed
Summary of Contents for Drexelbrook IntelliPoint RF RNL Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 12: ...Section 2...
Page 37: ...Troubleshooting 31 Section 5...
Page 41: ...Control Drawings 35 6 1 FM FMc Control Drawings Continued...
Page 42: ...RNLXX1 IntelliPoint RF Series User s Manual 36 6 1 FM FMc Control Drawings Continued...
Page 43: ...Control Drawings 37 6 1 FM FMc Control Drawings Continued...
Page 44: ...RNLXX1 IntelliPoint RF Series User s Manual 38 6 1 FM FMc Control Drawings Continued...
Page 45: ...Control Drawings 39 6 1 FM FMc Control Drawings Continued...
Page 46: ...RNLXX1 IntelliPoint RF Series User s Manual 40 6 1 FM FMc Control Drawings Continued...
Page 54: ...RNLXX1 IntelliPoint RF Series User s Manual 48 6 2 ATEX Control Drawings...
Page 55: ...Control Drawings 49 6 2 ATEX Drawings Continued...
Page 63: ...Control Drawings 57 6 3 CE Declaration of Conformity...
Page 64: ...RNLXX1 IntelliPoint RF Series User s Manual 58 6 4 IECEx Control Drawings...
Page 65: ...Control Drawings 59 6 4 IECEx Control Drawings Continued...
Page 73: ...Control Drawings 67 6 5 Mounting and Wiring for Spark Protector Drawings...
Page 75: ...Control Drawings 69 6 5 Mounting and Wiring for Spark Protector Continued...
Page 76: ...RNLXX1 IntelliPoint RF Series User s Manual 70 6 6 Adding a Padded Capacitor...
Page 77: ...Control Drawings 71 6 6 Adding a Padded Capacitor Continued...
Page 78: ...RNLXX1 IntelliPoint RF Series User s Manual 72 6 6 Adding a Padded Capacitor Continued...
Page 79: ...Control Drawings 73 6 7 Dual Seal Assembly for 700 Series Sensing Elements...
Page 80: ...RNLXX1 IntelliPoint RF Series User s Manual 74 6 7 Dual Seal Assembly Continued...
Page 81: ...Control Drawings 75 6 7 Dual Seal Assembly Continued...
Page 82: ...Appendix A...
Page 88: ......