Fields:
•
Location Name - Enter the location ID.
•
Sample Every - Available values are 1, 2, 5, 10, 15, 30, 60, and 120 minutes.
Aim to keep 651 samples < 150 readings per regeneration.
Aim to keep 451 samples < 100 readings per regeneration.
WARNING!
Sample intervals less than 15 minutes will significantly reduce the life of
the gold film sensor in the Jerome 631-X (or 431-X).
•
Regenerate Sensor Every - Available values are 6, 12, 24, and 48 hours. The default
is 48 hours.
An interval less than 48 hours will shorten the life of the gold film sensor in the 631-X
(or 431-X).
•
First Regen At - Regeneration takes an hour (no sampling), so pick a time when
sampling is not required. If regen interval is 12 hours and first regen is 12AM, then
the instrument will regen at 12AM and 12PM. For H2S odor control, midnight may be
a good time because it's cooler, so H2S is typically lower.
•
Enable Polling - This is usually checked. Leave it unchecked if this instrument is
offline.
•
Clear Data After Download - This should be checked. Typically data is deleted after
successful polling. If unchecked, the instrument will store readings until full (50k+
readings), which will take excessive time to download during polling. For sites with
difficult radio topology, transmission time must be minimized.
•
Double-Check High Readings Can Be Configured at the Instrument - This is set on
the instruments SAMPLE / DBL CHK menu.
•
Radio Address - Each instrument must have a unique radio address from 2 to 254.
The address in JMS must match the instrument. For the Nano radio, the base station
radio address is 1. See the instrument SYSTEM / RADIO menu (
to
program the instrument radio.
•
Pin Designation - A 1 or 2 digit designation that is used for the map display. If specified,
then the designation prefixes the instrument location name in the instruments list.
•
Instrument Is Also a Repeater - Instrument radios can be set as a repeater to other
instruments that don't have a line of site to the base station radio. Instrument radio
repeater settings are set at the instrument. This JMS field is only for informational
purposes and has not functional impact.
•
[Reset Location Information] - Use this button to clear the current instrument pin
location on the Site View tab.