7.2.5 Frequency sweep transient
Frequency step transients will cause the output frequency to go to the
value specified in the Goto F field starting at the phase angle specified in the
phase ø field for a duration specified in the Delay (s) field. The delay value
can range from 1 ms to 9999 seconds. At the end of the delay period, the
output frequency will remain at the Goto F value so there is no need to
specify a Final V value.
56
California Instruments
Rev D – March 2011
The illustration shown here reflects an example of a typical F Step
transient. The required row in the data entry screen for this transient would
be as shown below. In this case, register 0 was used to store the transient
information.
Frequency Step to 100 Hz for 120 ms.
120 ms