background image

 

 2  

098-0003 Rev. B 

 

ENDOTOXIN FILTER: 
 

The Endotoxin-Retentive Filters are the final purification process prior to distribution of the 
product water.  They are sub-micron filters with the ability to remove bacteria and endotoxin.  
The filters are enclosed in an opaque housing to inhibit the proliferation of algae.  Water flows 
through the filter, which traps bacteria and endotoxin. 
 
The MOBILE SPD SYSTEM is shipped completely assembled with required water treatment 
components.  This Operation Manual was written specifically for the MOBILE SPD SYSTEM.  
Your system was thoroughly tested and in excellent condition when it was shipped to you.  
However, because damage during shipment is possible, please unpack and carefully inspect 
the system as soon as you receive it.  Please notify AmeriWater if any problems are 
encountered. 
 

Please read the Operations Manual before using the system.

  Contact AmeriWater 

Customer Service with any questions at 1-800-535-5585 Monday through Friday 8:00 a.m. to 
5:00 p.m. eastern standard time.  For after hours emergencies call 1-800-535-5585 and 
follow the instructions on the recorded message.  Our on-call technician will return your call 
as soon as possible.  This entire Operations Manual should be read before operating or 
servicing the system.  This Operations Manual should then be kept near the system and used 
as a reference and troubleshooting guide. 

 
 1.1     SPECIFICATIONS 

 

Connections 

 

SRO Feed = 1/2" O.D. Tube 

SRO Product  = 1/2" O.D. Tube 

SRO Reject to Drain = 1/2" O.D. Tube 

 

Wall Mount DI Feed = 1/2” O.D. Tube 

W

all Mount DI Product = 1/2” MPT to 3/4” FGHT 

 

Storage Tank Feed = 3/4” Male GHT 

Storage Tank Product = 1

” FPT 

Storage Tank Return = 3/4” FPT 

 

Pump Suction = 1” FPT 

Pump Discharge = 

1” FPT 

 

Endotoxin Filter Feed = 3/4”FPT 

En

dotoxin Product = 3/4”FPT 

 

 

Electrical Requirements 

 

 

3X 120V 1PH Dedicated 20Amp GFI service 

 

Shipping Weight 

Operating Weight 

353 LBS 
796 LBS 

Summary of Contents for SPD

Page 1: ...098 0003 Rev B MOBILE SPD FOR STERIS OPERATION MAINTENANCE MANUAL Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414...

Page 2: ...098 0003 Rev B...

Page 3: ...RAM 4 2 0 SYSTEM INSTALLATION 5 2 1 SRO 5 2 2 WALL MOUNT DI 6 2 3 STORAGE TANK 8 2 4 ENDOTOXIN FILTER 9 3 0 SYSTEM START UP 9 3 1 OPERATION 10 4 0 MONITORING 11 5 0 PREVENTATIVE MAINTENANCE SCHEDULE F...

Page 4: ...cation resin bead then an anion then another cation another anion and so on An exchange takes place of course only when a positive ion contacts a negative exchange site and vice versa With each excha...

Page 5: ...he Operations Manual before using the system Contact AmeriWater Customer Service with any questions at 1 800 535 5585 Monday through Friday 8 00 a m to 5 00 p m eastern standard time For after hours e...

Page 6: ...to electrical requirements system pressures and voltage draws please refer to listed component manuals as listed below For 00SRO1095 RO refer to Manual 98 0150 For 00MW200 Silex Wall Mount Deionizer...

Page 7: ...4 098 0003 Rev B 1 2 SYSTEM FLOW DIAGRAM Figure 1 1...

Page 8: ...e wall with anchoring bolts 3 SRO is equipped with sediment and chlorine removal 4 Connect the feed water supply to the inlet 1 2 fitting on the SRO prefilter 5 The feed water supply requires a minimu...

Page 9: ...e of supporting 50 pounds 1 Locate the DI Polisher to the right of the SRO just before the storage tank and within 6 feet from a 120 volt 60 hertz receptacle 2 Assembly should be level when mounted to...

Page 10: ...1 2 MPT x 3 4 GHT outlet fitting Figure 2 2 FIGURE 2 2 5 Make sure that the deionizing cartridges are installed in the filter housing and that the bowls are hand tightened only 6 Plug transformer into...

Page 11: ...ct the Wall Mount DI product outlet hose to the storage tank s inlet Figure 2 2 above 3 Connect the tank outlet to the distribution pump Figure 2 3 FIGURE 2 3 4 Run the float switch quick disconnect f...

Page 12: ...to prevent microbiological growth and freezing Discard all product water for at least two hours of operation before placing the RO in service NOTE As each component is started and flushed of any debri...

Page 13: ...at the TDS is above the pre determined set point the alarm lamp will light and the HI TDS warning message will show on the display This warning will clear when the TDS drops below the Setpoint The DI...

Page 14: ...uality management system using continuous reading monitors that measure product water quality total dissolved solids These measurements can provide a measure of equipment performance 3 Cartridge filte...

Page 15: ...PARAMETERS INITIALS Feed Water Temperature 50 F 80 F Feed Water Pressure 20PSI Pump PSI 100 200PSI Product Flow GPM 0 55 0 76 Reject Flow GPM ADJUSTED EQUAL TO PRODUCT Product Quality TDS 25ppm Total...

Page 16: ...r 20 6001 must be replaced in the event of listed instances below 1 Red light the red light warns that the water conductivity is above the threshold value of 200 000 ohm For 0185 8000 Storage Tank Ven...

Page 17: ...rees F Gages are located on face plate of the RO and top of the Pre filter housing Return on the Storage Tank and the inlet and outlet of Endotoxin filter Fig 5 1 Step 4 Membranes Remove membranes fro...

Page 18: ...15 098 0003 Rev B Figure 5 1...

Page 19: ...0024 END CAP 2 25 and 40 SS HSG SMALL BLUE O RING RO Membrane 21 0035 O RING FLTR HOUS 10 SLIMLINE BLUE BLK 50I O NO PR Prefilter Vent Filter and DI 21530233 O RING FLTR HSG 10IN BB SQ THD Loop Endoto...

Page 20: ...ifornia to cause cancer birth defects or other reproductive harm One such chemical is Vinyl Chloride a compound used to produce Polyvinyl Chloride PVC The AmeriWater system you have purchased may cont...

Reviews: