5.1  

WINTERIZATION  

 
Step 1 

– 

 
Disconnect all Power sources. 
 
Filter housings- Open the sample ports to all filter housings in the system (i.e. RO prefilter, 
Wall Mount DI, Endotoxin filters) to relieve water pressure. Open filter housings, remove 
filters cartridges and discard cartridges. Undo any plumbing connections to the filter housing 
that will not allow water to freely drain from the fitting. 
 
 
Step 2 

– 

 
Drain water from system. Open all plugs. Disconnect all joints.  Save all plugs and O-rings / 
gaskets. (Fig. 5.1) 
 
 
 
Step 3 

– 

 
Gages 

– Because gages are liquid filled.  Remove from RO, Storage Tank and from 

Endotoxin Filter. Store in an area above 34 degrees F. Gages are located on face plate of the 
RO and top of the Pre filter housing, Return on the Storage Tank and the inlet and outlet of 
Endotoxin filter.  (Fig. 5.1) 
 
 
Step 4 

–  

 
Membranes 

– Remove membranes from housings and discard. Disconnect joints to 

membrane. Remove all water pooling in any joints. Remove any connection that will not allow 
water to free flow out of it.  Uncouple the product line and drain all water. Remove any fitting 
that will not allow water to flow freely out of it. (Fig. 5.1) 
 
Step 5 

– 

 
RO Pump Assembly 

– Disconnect all fitting to the pump assembly. Allow pump housing to 

completely drain. 
 
Step 6 

– 

 
Demand pump 

– MQ pump. Remove drain cock on the MQ pump and allow to drain. 

 
 
Step 7 

– 

 

Wall Mount DI 

– Open DI housings, remove DI cartridges and discard cartridges. Completely 

drain housings. Once housings are drained, put housing O-rings in housing bowl. 

Summary of Contents for SPD

Page 1: ...098 0003 Rev B MOBILE SPD FOR STERIS OPERATION MAINTENANCE MANUAL Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414...

Page 2: ...098 0003 Rev B...

Page 3: ...RAM 4 2 0 SYSTEM INSTALLATION 5 2 1 SRO 5 2 2 WALL MOUNT DI 6 2 3 STORAGE TANK 8 2 4 ENDOTOXIN FILTER 9 3 0 SYSTEM START UP 9 3 1 OPERATION 10 4 0 MONITORING 11 5 0 PREVENTATIVE MAINTENANCE SCHEDULE F...

Page 4: ...cation resin bead then an anion then another cation another anion and so on An exchange takes place of course only when a positive ion contacts a negative exchange site and vice versa With each excha...

Page 5: ...he Operations Manual before using the system Contact AmeriWater Customer Service with any questions at 1 800 535 5585 Monday through Friday 8 00 a m to 5 00 p m eastern standard time For after hours e...

Page 6: ...to electrical requirements system pressures and voltage draws please refer to listed component manuals as listed below For 00SRO1095 RO refer to Manual 98 0150 For 00MW200 Silex Wall Mount Deionizer...

Page 7: ...4 098 0003 Rev B 1 2 SYSTEM FLOW DIAGRAM Figure 1 1...

Page 8: ...e wall with anchoring bolts 3 SRO is equipped with sediment and chlorine removal 4 Connect the feed water supply to the inlet 1 2 fitting on the SRO prefilter 5 The feed water supply requires a minimu...

Page 9: ...e of supporting 50 pounds 1 Locate the DI Polisher to the right of the SRO just before the storage tank and within 6 feet from a 120 volt 60 hertz receptacle 2 Assembly should be level when mounted to...

Page 10: ...1 2 MPT x 3 4 GHT outlet fitting Figure 2 2 FIGURE 2 2 5 Make sure that the deionizing cartridges are installed in the filter housing and that the bowls are hand tightened only 6 Plug transformer into...

Page 11: ...ct the Wall Mount DI product outlet hose to the storage tank s inlet Figure 2 2 above 3 Connect the tank outlet to the distribution pump Figure 2 3 FIGURE 2 3 4 Run the float switch quick disconnect f...

Page 12: ...to prevent microbiological growth and freezing Discard all product water for at least two hours of operation before placing the RO in service NOTE As each component is started and flushed of any debri...

Page 13: ...at the TDS is above the pre determined set point the alarm lamp will light and the HI TDS warning message will show on the display This warning will clear when the TDS drops below the Setpoint The DI...

Page 14: ...uality management system using continuous reading monitors that measure product water quality total dissolved solids These measurements can provide a measure of equipment performance 3 Cartridge filte...

Page 15: ...PARAMETERS INITIALS Feed Water Temperature 50 F 80 F Feed Water Pressure 20PSI Pump PSI 100 200PSI Product Flow GPM 0 55 0 76 Reject Flow GPM ADJUSTED EQUAL TO PRODUCT Product Quality TDS 25ppm Total...

Page 16: ...r 20 6001 must be replaced in the event of listed instances below 1 Red light the red light warns that the water conductivity is above the threshold value of 200 000 ohm For 0185 8000 Storage Tank Ven...

Page 17: ...rees F Gages are located on face plate of the RO and top of the Pre filter housing Return on the Storage Tank and the inlet and outlet of Endotoxin filter Fig 5 1 Step 4 Membranes Remove membranes fro...

Page 18: ...15 098 0003 Rev B Figure 5 1...

Page 19: ...0024 END CAP 2 25 and 40 SS HSG SMALL BLUE O RING RO Membrane 21 0035 O RING FLTR HOUS 10 SLIMLINE BLUE BLK 50I O NO PR Prefilter Vent Filter and DI 21530233 O RING FLTR HSG 10IN BB SQ THD Loop Endoto...

Page 20: ...ifornia to cause cancer birth defects or other reproductive harm One such chemical is Vinyl Chloride a compound used to produce Polyvinyl Chloride PVC The AmeriWater system you have purchased may cont...

Reviews: