Fill out the monitoring log daily.  Having this information available will help to quickly diagnose 
issues related to performance.  Failure to carry out the daily monitoring and maintenance at 
the indicated intervals will result in reduced performance of the SRO system and may void 
the warranty. 
 

(SHOULD BE COMPLETED DAILY WHEN THE SYSTEM IS USED) 

DATE

 

 

 

 

SRO PERFORMANCE 

     PARAMETERS 

RECORDED 
PARAMETERS 

INITIALS 

    Feed Water Temperature

 

       50ºF - 80ºF

 

 

 

    Feed Water Pressure

 

         >20PSI 

 

 

    Pump PSI

 

       

100 - 200PSI 

 

 

    Product Flow - GPM

 

        

0.55 - 0.76 

 

 

    Reject Flow   -  GPM

 

ADJUSTED EQUAL   

 TO PRODUCT

 

 

 

   Product Quality - TDS

 

        < 25ppm 

 

 

   Total Chlorine 

       < 0.1ppm 

 

 

   Reject PSI 

       < 135PSI 

 

 

WALL MOUNT DEIONIZER 

 

 

 

   Product Quality >100,000OHM when  

   operating 

      GREEN LIGHT 

 

 

STORAGE TANK 

 

 

 

   Loop Return PSI 

          

>30 PSI 

 

 

   Vent Filter Dry 

         YES/NO 

 

 

DISTRIBUTION PUMP 

 

 

 

   Pump PSI  

             >65psi 

 

 

ULTRA FILTER 

 

 

 

   Filter 

ΔP 

          <15PSI 

 

 

EXCHANGE

 

 

 

 

    10 Micron Carbon Block Pre-Filter 

      AS NEEDED 

 

 

    Membrane

 

      AS NEEDED

 

 

 

    Deionizer Cartridges

 

      RED LIGHT

 

 

 

    Storage Tank Vent Filter

 

    YEARLY / WET

 

 

 

    Endotoxin Filter Cartridges 

           

>15 ΔP

 

 

 

 

 

Summary of Contents for SPD

Page 1: ...098 0003 Rev B MOBILE SPD FOR STERIS OPERATION MAINTENANCE MANUAL Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414...

Page 2: ...098 0003 Rev B...

Page 3: ...RAM 4 2 0 SYSTEM INSTALLATION 5 2 1 SRO 5 2 2 WALL MOUNT DI 6 2 3 STORAGE TANK 8 2 4 ENDOTOXIN FILTER 9 3 0 SYSTEM START UP 9 3 1 OPERATION 10 4 0 MONITORING 11 5 0 PREVENTATIVE MAINTENANCE SCHEDULE F...

Page 4: ...cation resin bead then an anion then another cation another anion and so on An exchange takes place of course only when a positive ion contacts a negative exchange site and vice versa With each excha...

Page 5: ...he Operations Manual before using the system Contact AmeriWater Customer Service with any questions at 1 800 535 5585 Monday through Friday 8 00 a m to 5 00 p m eastern standard time For after hours e...

Page 6: ...to electrical requirements system pressures and voltage draws please refer to listed component manuals as listed below For 00SRO1095 RO refer to Manual 98 0150 For 00MW200 Silex Wall Mount Deionizer...

Page 7: ...4 098 0003 Rev B 1 2 SYSTEM FLOW DIAGRAM Figure 1 1...

Page 8: ...e wall with anchoring bolts 3 SRO is equipped with sediment and chlorine removal 4 Connect the feed water supply to the inlet 1 2 fitting on the SRO prefilter 5 The feed water supply requires a minimu...

Page 9: ...e of supporting 50 pounds 1 Locate the DI Polisher to the right of the SRO just before the storage tank and within 6 feet from a 120 volt 60 hertz receptacle 2 Assembly should be level when mounted to...

Page 10: ...1 2 MPT x 3 4 GHT outlet fitting Figure 2 2 FIGURE 2 2 5 Make sure that the deionizing cartridges are installed in the filter housing and that the bowls are hand tightened only 6 Plug transformer into...

Page 11: ...ct the Wall Mount DI product outlet hose to the storage tank s inlet Figure 2 2 above 3 Connect the tank outlet to the distribution pump Figure 2 3 FIGURE 2 3 4 Run the float switch quick disconnect f...

Page 12: ...to prevent microbiological growth and freezing Discard all product water for at least two hours of operation before placing the RO in service NOTE As each component is started and flushed of any debri...

Page 13: ...at the TDS is above the pre determined set point the alarm lamp will light and the HI TDS warning message will show on the display This warning will clear when the TDS drops below the Setpoint The DI...

Page 14: ...uality management system using continuous reading monitors that measure product water quality total dissolved solids These measurements can provide a measure of equipment performance 3 Cartridge filte...

Page 15: ...PARAMETERS INITIALS Feed Water Temperature 50 F 80 F Feed Water Pressure 20PSI Pump PSI 100 200PSI Product Flow GPM 0 55 0 76 Reject Flow GPM ADJUSTED EQUAL TO PRODUCT Product Quality TDS 25ppm Total...

Page 16: ...r 20 6001 must be replaced in the event of listed instances below 1 Red light the red light warns that the water conductivity is above the threshold value of 200 000 ohm For 0185 8000 Storage Tank Ven...

Page 17: ...rees F Gages are located on face plate of the RO and top of the Pre filter housing Return on the Storage Tank and the inlet and outlet of Endotoxin filter Fig 5 1 Step 4 Membranes Remove membranes fro...

Page 18: ...15 098 0003 Rev B Figure 5 1...

Page 19: ...0024 END CAP 2 25 and 40 SS HSG SMALL BLUE O RING RO Membrane 21 0035 O RING FLTR HOUS 10 SLIMLINE BLUE BLK 50I O NO PR Prefilter Vent Filter and DI 21530233 O RING FLTR HSG 10IN BB SQ THD Loop Endoto...

Page 20: ...ifornia to cause cancer birth defects or other reproductive harm One such chemical is Vinyl Chloride a compound used to produce Polyvinyl Chloride PVC The AmeriWater system you have purchased may cont...

Reviews: