PROCESS DESCRIPTION
This section provides general information regarding the heat disinfection
process utilized by the
Heatsan
unit.
Operation
General philosophy of operation:
The unit can be configured to disinfect up to two distribution loops,
either individually or together. The user can set for each day of the
week which loop is to be disinfected and in what configuration.
The unit has been designed to provide up to 95 gallons of hot water for
the purposes of sanitizing connected dialysis equipment.
The number of machines that can be disinfected at any one time will
be dependent on the requirement of those machines regarding the
amount of water each machine needs for disinfection, the time taken
to perform heat disinfection and the flow at which they require the hot
water.
The system allows the user to input the specific requirements of the
dialysis machines to be disinfected, the days requiring heat disinfection
and the times which heat disinfection can occur.
If the system detects that any of the entered user values cannot be
achieved by the Heatsan unit the unit will raise an
“Invalid number”
alarm warning the user that the parameters loaded cannot be
achieved.
The typical stage duration times for disinfection can be found in
Section 6.10 Factory Default Settings.
Note: These values can be adjusted to fit requirements of your
particular distribution system. Verification of the efficacy of changes is
the responsibility of the medical director.
Preheat
The preheat stage is initiated prior to the Heatsan start time. The
start and return to service times being set by the clinical staff.
Water is taken off of the loop during the preheat stage.
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...