38
98-0156 Rev G
Factory Default Settings
Refer to Section 6.6 Factory set-up to change and save default values.
Parameter
Unit
Factory Default
Values
Screen Location
Number of Machines
Number
1
Current Setup
Heatsan Start
Time
0:00
Current Setup
Service Start
Time
0:00
Current Setup
Maximum Dialysis Machines
Number
17
System Setup Page 1
Sanitization Temp
o
F
185
System Setup Page 1
Tank Pre-Heat Temp
o
F
150
System Setup Page 1
Tank Pre-Heat Time
Hr/min
2:00
System Setup Page 2
Loop Pre-Heat Time
Hr/min
2:00
System Setup Page 2
Hold Time
Hr/min
0:30
System Setup Page 2
Draw Off Time
Hr/min
0:30
System Setup Page 2
Cool Down Time
Hr/min
1:00
System Setup Page 2
Maximum Operating Temp
o
F
197
Factory Setup Page 1
Minimum Operating Temp
o
F
50
Factory Setup Page 1
Maximum Service Temp
o
F
88
Factory Setup Page 1
Minimum Hold Time
Hr/min
0:30
Factory Setup Page 1
Maximum Draw Off Available
USGal
Not Adjustable
Factory Setup Page 2
Draw Off
Gal/Mach 3
Factory Setup Page 2
Rate
GPM
1
Factory Setup Page 2
Time per Dialysis Machine
Hr/min
0:15
Factory Setup Page 3
Water Loss in Hold
UsGal
5
Factory Setup Page 3
Heater 1 Level
UsGal
15
Factory Setup Page 3
Heater 2 Level
UsGal
21
Factory Setup Page 3
Heater 3 Level
UsGal
32
Factory Setup Page 3
Cool Down Temperature Drop
o
F
8
Factory Setup Page 4
(Upper Limit) Temp Hysteresis
o
F
2
Factory Setup Page 4
(Lower Limit) Heat Loss
o
F
7
Factory Setup Page 4
Maximum Tank Level
UsGal
Not Adjustable
Factory Setup Page 4
Minimum Tank Level
UsGal
Not Adjustable
Factory Setup Page 4
Water Feed On
o
F
189
Factory Setup Page 5
Water Feed Off
o
F
184
Factory Setup Page 5
Refill Period Limit
Hr/min
0:30
Factory Setup Page 5
Draw Off End Warning
Secs
Display only
Factory Setup Page 5
Note:
After making any changes to the information in the Factory Setup screens
save it by accessing the “Save Defaults” and pressing the button for a
minimum of 5 seconds or until you hear a confirmation “bleep”.
Summary of Contents for Heatsan 00HS208
Page 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Page 52: ...51 98 0156 Rev G Wiring Diagrams ...
Page 53: ...52 98 0156 Rev G ...
Page 54: ...53 98 0156 Rev G ...
Page 55: ...54 98 0156 Rev G ...
Page 56: ...55 98 0156 Rev G ...
Page 57: ...56 98 0156 Rev G ...
Page 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Page 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Page 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Page 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Page 70: ...69 98 0156 Rev G ...