27
Alarm Relay Trip Point Selection Table
Volts PPM PPM PPM PPM PPM PPM
PPM
PPM PPM PPM
PPM
100% Full
Scale
2.0 1 3 10 25 50 100
200
300
400
500 999
75% Full
Scale
1.6 0.75 2.25 7.5 18.75
37.5 75 150 225 300 375 749.25
50% Full
Scale
1.2 0.5 1.5 5 12.5 25 50 100 150 200 250 499.5
25% Full
Scale
0.8 0.25 0.75 2.5 6.25 12.5 25 50 75 100 125 249.75
0% Full
Scale
0.4
0 0 0 0 0 0 0 0 0 0 0
Alarm Relay Trip Point Selection
10 PPM
25 PPM
50 PPM
100 PPM
200 PPM
300 PPM
400 PPM
500 PPM
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
2.0
0
20
40
60
80
100
120
140
160
180
200
220
240
260
280
300
320
340
360
380
400
420
440
460
480
500
PPM
V
o
ltag
e (
V
)
100
75 %
50 %
25 %
Figure 5-2 Alarm Relay Trip Point Section Graph
Summary of Contents for VersaTox