14
Neonate
Pediatric
Adult
Inspiratory resistance
***
,
*****
with oxygen
reservoir tube:
0�6 cm H₂O
at 5 l/min
4�6 cmH₂O
at 50 l/min
with oxygen
reservoir bag:
0�9 cm H₂O
at 5 l/min
5�0 cm H₂O
at 50 l/min
with oxygen
reservoir bag:
0�7 cmH₂O
at 5 l/min
5�0 cmH₂O
at 50 l/min
with oxygen
reservoir bag:
3�7 cmH₂O
at 50 l/min
Expiratory resistance
***
,
*****
1�6 cm H₂O
at 5 l/min
4�5 cmH₂O
at 50 l/min
2�3 cmH₂O
at 50 l/min
PEEP generated by resuscitator
in normal use with added supply
gas flow
***
,
*****
with oxygen
reservoir tube:
< 2 cm H₂O
at 5, 10,
and 15 l/min
with oxygen
reservoir bag:
2�5 cm H₂O
at 5 l/min
2�8 cm H₂O
at 10 l/min
2�8 cm H₂O
at 15 l/min
(V
T
20 ml, f 60)
with oxygen
reservoir bag:
< 2 cm H₂O
at 5, 10,
and 15 l/min
(V
T
150 ml, f 25)
with oxygen
reservoir bag:
< 2 cm H₂O
at 5, 10,
and 15 l/min
(V
T
450 ml, f 20)
Neonate
Pediatric
Adult
Patient connector
Outside 22 mm male (EN ISO 5356-1)
Inside 15 mm female (EN ISO 5356-1)
Expiration connector
(for PEEP valve attachment)
30 mm male (EN ISO 5356-1)
Manometer port connector
Ø 4�2 +/- 0�1 mm
Bag refill valve connector
*
-
Inside 32 mm
female
Inside 32 mm
female
Forward and backward leakage
Not measurable
O₂ inlet connector
According to EN ISO 13544-2
Operation temperature limits
*
-18 °C to +50 °C (-0�4 °F to +122 °F)
Storage temperature limits
*
-40 °C to +60 °C (-40 °F to +140 °F)
Recommended long term storage in closed packaging at room temperature,
away from sunlight�
NOTES:
– 10 cmH₂O = 1�0 kPa
– V
T
: Ventilation volume
– f: Frequency (breath per minute)
*
Tested according to EN ISO 10651-4
**
Higher airway pressure can be obtained by overriding the pressure-limiting valve
***
At general test conditions according to EN ISO 10651-4:2009
****
Values are approximate
*****
Maximum values