
en
11
• If continued resistance to insufflation is encountered, check the airway for obstruction
and re-position the patient, to ensure an open airway�
• If the patient vomits during ventilation; immediately remove the resuscitator to
clear the patient's airway and expel the vomit from the resuscitator by shaking and
compressing it forcefully and fast several times before resuming ventilation� In the
case of excessive amount of vomitus obstructing the air flow, the patient valve may be
disassembled and cleaned� For details on dis- and reassembly of the patient valve refer
to illustrations
5.5
and
6.1
�
• If connecting external devices to the resuscitator, make sure to test for functionality and
consult the instructions for use accompanying these external devices�
Manometer port
The Ambu Disposable Pressure Manometer or a third-party pressure gauge can be
attached to the manometer port, situated on the top of the patient valve� Remove the cap
and attach the manometer/pressure gauge
8
�
Pressure-limiting valve
The pressure-limiting valve is set to open at 40 cmH₂O (4�0 kPa)�
If medical and professional assessment indicates that a pressure above 40 cmH₂O is
required, the pressure-limiting valve can be overridden by pressing the override cap onto
the valve
3.2
�
Alternatively, the pressure-limiting valve can be overridden by placing the index finger on
the blue button while squeezing the bag�
Oxygen administration
Administer oxygen according to the medical indication�
Figure 4 shows calculated delivered oxygen percentages which can be obtained with
different ventilation volumes and frequencies at different gas flow rates referring to Oval
Plus Silicone Resuscitator Adult
4.1
, Oval Plus Silicone Resuscitator Pediatric
4.2
and Oval
Plus Silicone Resuscitator Neonate
4.3
, respectively�
4.4. Reprocessing: cleaning, disinfection, sterilization
Follow these reprocessing instructions after each use to reduce the risk of cross-contamination�
Disassembly
Before reprocessing, manually disassemble the resuscitator into individual components to
the level shown in
5.1
5.2
5.3
to make surfaces accessible to cleaning�
Follow the disassembly method shown in
5.4
5.5
and
5.6
�
Keep components from the same device together during reprocessing and keep track
of the number of times each resuscitator has been processed to avoid reassembly of
components with different durability�
Recommended reprocessing procedures
For a complete reprocessing of the Ambu Oval Plus Silicone Resuscitator, use one of the
procedures listed in Table 1�
Product/Component
Recommended reprocessing procedures (select one)
• Oval Plus Silicone
Resuscitator Adult
• Oval Plus Silicone
Resuscitator Pediatric
• Oval Plus Silicone
Resuscitator Neonate
(except oxygen
reservoir tube)
• Manual cleaning followed by chemical disinfection�
• Manual cleaning followed by sterilization�
• Automated cleaning, including a thermal disinfection
stage, followed by sterilization�
• Automated cleaning, including a thermal disinfection
stage, followed by chemical disinfection�
Oxygen reservoir tube
for Oval Plus Silicone
Resuscitator Neonate
• Manual cleaning followed by chemical disinfection�
Table 1: Recommended reprocessing procedures.