5.21
Stop at zero
Section:
ConF
Display:
StOP ZERO
Range:
OFF
or
ON
Default Value:
OFF
Default Access Level
4
Function number
4C98
This function is only available when the down timer mode is selected at the
ConF OPEr
function. When set to
ON
the down count will stop at zero. When set to
OFF
the timer will continue through zero to show negative value
if the timer operation is still active.
5.22
Display running indication
Section:
ConF
Display:
RUN FLSH
Range:
OFF
or
ON
Default Value:
On
Default Access Level
4
Function number
4C9A
The colons and/or decimal point depending on display type on the display can be set to flash when the display is
running (timing up or down) (
RUN FLSH
set to
ON
) or can be set not to flash at any time (
RUN FLSH
set to
OFF
). The flashing mode is useful to indicate the display is running when the display is set for minutes, hours or
days.
5.23
Reset at power up
Section:
ConF
Display:
P.ON RSt
Range:
OFF
or
ON
Default Value:
OFF
Default Access Level
4
Function number
4C9b
When set to
On
the display will reset to zero each time it is powered up. When set to
OFF
the display will retain
and display its last value when powered up.
5.24
Inputs 1 and 2 pull up
Section:
IN 1
to
IN 2
Display:
PULL UP
Range:
OPEn
,
High
,
Lo
Default Value:
High
Default Access Level
4
Function number
4C00
to
4C01
Allows electrical input setting of inputs 1 and 2. When set to
High
approximately 5VDC will be placed on the
input internally. When set to
Lo
the input will be pulled down to 0V (signal ground). When set to
OPEn
signals
up to 48VDC can be applied. See “Electrical Installation” chapter for further information.
LD5TMRMAN-1.1-0
39 of