24
Eco/Classic Installation Guide and Technical Specification
3.6
Electric Emergency Lowering
The power supply unit fitted on the bed frame is equipped with a 9V block battery, which
makes it possible to make a CPR emergency lowering according to EN 60601-2-52 in
the event of a power failure. Please note, however, that this is only possible once per 9V
battery, as the capacity of the 9V battery is limited.
After the emergency lowering has been used once, the 9V battery must be replaced with
a new one (Type 6LR61 alkaline manganese battery). The 9V battery should however be
replaced every 2 years even if it has not been used.
3.7
9V Battery Change
To replace, check, or remove the 9V battery, open the battery compartment on the
power supply unit attached to the backrest motor.
Carry out the battery change as follows:
ATTENTION:
Unplug the mains plug!
• Take off the plug securing device by removing the two crosshead screws
• Now you may pull out the cover with the battery and replace it with the new one.
Connect the new battery and slide it in with the cover again.